Video summary

CHEM1811 Calibration of pipette and measuring cylinder (updated)

Main summary

Key takeaways

Educational

Main Ideas

Purpose of the Lab

The experiment calibrates a pipette and a measuring cylinder by comparing the volume each delivers with a volume calculated from the mass of water delivered and water’s density at the measured temperature. The goal is to assess the glassware’s accuracy and precision.

Types of Measurement Error

  • Gross errors: Procedural mistakes, such as forgetting to record a value, guessing a result, reading a meniscus incorrectly, or using a pipette improperly. Reject and repeat measurements affected by these errors.
  • Systematic errors (bias or constant errors): Consistent errors associated with equipment or experimental design. Repeated measurements do not remove them; calibration helps identify them.
  • Random errors: Unpredictable fluctuations inherent in measurement. They cannot be eliminated, but repeated measurements under similar conditions can reduce their effect on the reported average.

Accuracy and Precision

  • Accuracy describes how close the mean measured value is to the accepted or expected value.
  • Precision describes how closely repeated measurements agree. A smaller spread indicates greater precision.
  • Accuracy and precision are independent: results can be accurate but imprecise, precise but inaccurate, both, or neither.
  • Standard deviation quantifies spread. A low standard deviation indicates better precision; a high one indicates poorer precision. The sample standard deviation uses the sample mean and divides by (n-1), while the population standard deviation uses the population mean and divides by (n).
  • Relative error is calculated as standard deviation ÷ mean × 100%. Lower relative error indicates better precision. The talk gives less than 5% as a general rule of thumb for very good precision.

Experimental Method

Preparation and Data Collection

  • The lab has data-collection and data-analysis stages. Record observations directly on the worksheet in the lab manual, in pen, not pencil.
  • Wear a lab coat, safety glasses, and closed, protective footwear. Sandals, Crocs, and other open footwear are not accepted.
  • Measure and record water temperature as directed on the worksheet. Because the thermometer is marked in one-degree units, estimate and report the temperature to one decimal place.
  • Use the water temperature to determine water’s density from the provided density–temperature information, interpolating from the graph when necessary. Density changes with temperature, so it is needed to convert measured mass to volume.
  • For each glassware type, record the mass of the empty, dry beaker. Dispense the specified volume of water, then weigh the beaker and water. Continue with successive additions of water—do not discard the water between readings—and record each mass. The two glassware types may be tested in either order.
  • Carry out repeated measurements, at least five as described in the talk, so that a mean and standard deviation can be calculated.

Using the Pipette and Reading Glassware

  • Use a pipette filler; never pipette by mouth. Squeezing and releasing the filler changes the pressure so liquid is drawn into the pipette.
  • Bring the meniscus to the pipette’s calibration mark and use the pipette correctly. Incorrect technique is a procedural error that should be corrected by repeating the measurement.
  • Read liquid volumes at eye level to avoid parallax. This applies to pipettes, burettes, measuring cylinders, volumetric flasks, and other analytical glassware.

Using the Analytical Balance

Distinguish between the top-loading balance, which typically reads to one or two decimal places and can handle larger masses, and the more sensitive analytical balance, which typically reads to four decimal places.

When using the analytical balance:

  • Close the protective doors or panes before recording a reading to prevent drafts from affecting it.
  • Tare or zero the balance as appropriate, and wait for a steady reading. Do not wait excessively if the reading continues to drift.
  • Avoid large loads, unnecessary vibration, drafts, and moving the balance.
  • Do not handle weigh vessels with bare hands; use a suitable lint-free material or another approved method.
  • Do not add or transfer solids or liquids while a container is on the balance. Remove it first.
  • Clean up spills promptly.

Data Analysis and Significant Figures

Calculate the mass of water by subtracting the relevant beaker mass:

  • For the first addition: mass of beaker plus water minus mass of empty beaker.
  • For each later addition: the new beaker-plus-water mass minus the previous beaker-plus-water mass. This is weighing by difference.

Then:

  1. Calculate the water volume using its mass and density at the measured temperature.
  2. Find the mean volume, each measurement’s deviation from the mean, and the squared deviations; use these to calculate the standard deviation.
  3. Apply significant-figure rules according to each operation: subtraction is governed by decimal places, while multiplication and division are governed by significant figures.

Report final results in a way that reflects the glassware’s intended capacity, rather than copying all calculator digits:

  • The example pipette is nominally 5.00 mL, reflecting its stated precision to two decimal places.
  • The measuring cylinder should be reported to one decimal place.
  • The standard deviation may be reported to one additional decimal place, depending on its size.

The worksheet supplies formulas for calculations not covered in the introductory material.

Possible Sources of Error

  • Gross: Parallax, failing to align the pipette meniscus with its calibration mark, or otherwise mishandling the glassware. Correct these errors by repeating the measurement rather than treating them as unavoidable.
  • Systematic: Possible equipment or method bias, including whether temperature is accounted for correctly.
  • Other experimental effects: Water may evaporate from open containers; consider whether evaporation could affect results.
  • Random: Measurement-to-measurement fluctuations; reduce their impact through repeated measurements.

Lab Requirements and Reminders

  • Bring the prelab graph of water temperature as a function of density; the instructor says it is required for prelab credit.
  • Complete the worksheet and report before the lab session ends. Use waiting time to record data and complete calculations rather than leaving work until the end.
  • Cell phones are not permitted in the lab space. The instructor warns that using one can result in a zero and that cheating materials will be dealt with seriously.

Speakers and Sources

  • Speaker: One unnamed instructor or narrator presenting the lab talk and explaining the procedures.
  • Other identifiable speakers: None. The subtitles mention brief pipette and balance demonstrations but do not identify any additional speakers.

Rate this summary

Your feedback will help improve summaries.

Improve this summary

Reprocess with a stronger model when the summary feels incomplete or inaccurate.

Pro

Translate summary in another language

Pro

Ask questions to this video

Chat for follow-up questions, clarifications, and source-backed answers.

Coming soon

Share this summary

Original video